We calibrate your station onsite to prevent long downtimes.
up to 70barg

Calibration of thermocouples up to 100°C

MFM and MFC calibration from 20Nml/min up to 8000Nl/min
MFCs will be calibrated in their installed position. MFCs will be calibrated with customer gas infrastructure, including its limitations. Available for H2, N2, O2, Air and CO2 (O2 and CO2 with Bronkhorst correction factors). With TIRRON calibration certificates.
Does not include repair of MFCs, which cannot be calibrated. (MFC can not be calibrated within our tolerances)
TIRRON Reference is calibrated at ISO 17025 certified laboratory with 1 year calibration interval. With 5 calibration points: 0, 20, 40, 60, 80 and 100% FS.
Tolerance and Uncertainty:
20nmlpm - 1000 nlpm:
-Tolerance: 1 % Measurement + 0.2 % Full Scale of DUT
-Uncertainty: 0.5 % Measurement + 0.1% Full Scale of used References

Load and Powersupply calibration up to 1000A

Resistivity validation

To keep your station safe and reliable
- safety matrix check
- leak check
- pump maintenance
- filter maintenance
We create calibration certificates for all calibrates devices.